Oasis Instrument Co., Ltd. is experienced in the design, manufacture, and commissioning of advanced water flow calibration systems. Our expertise enables us to create systems with pipe diameters ranging from DN15 to DN1200. These calibration systems are suitable for verifying a variety of flowmeter types, including velocity-type flowmeters, positive displacement flowmeters, Coriolis mass flowmeters, and differential pressure flowmeters. Our systems support real-time data acquisition, automatic data processing, automatic calibration certificate generation, and historical data retrieval.
Main Technical Specifications of the Systems:
Verification Methods: Static mass method and Master flowmeter method.
Pipe Diameter Range: DN15 to DN1200.
Flow Range: Customized according to client requirements.
Static Gravimetric Method: 0.2m3/h~5000 m3/h
Master Meter: 0.2m3/h~8500 m3/h
Static Volumetric Method: 0.2m3/h~1530 m3/h
Device Uncertainty (K=2):
- Static mass method: ≤0.05%
- Standard flowmeter method: ≤0.2%
- Static Volume method: ≤0.05%
Minimum Measurement Time: 30 seconds.
Design Medium Temperature Range: 5 to 35 °C.
Flow Stability: ≤ 0.15% for the water flow calibration device.
Working Pressure (Upstream Pressure of Flowmeter): 0.2 to 0.4 MPa.
Design Pressure in Pipeline: 1.6 MPa.
Calibration Medium: Water.